Mortara ELI250c ECG Machine
The ELI™ 250c provides full-featured functionality in a simple-to-learn and easy-to-use portable unit.
The ELI 250c offers Best 10, full size or A4 paper, and the option of traditional or wireless ECG acquisition.
This product is currently on offer with NHS Supply Chain. Click here for full details.
Product Features
Portable Solution
Compact and lightweight, the ELI™ 250c electrocardiograph
provides comprehensive functionality, with 8.5 x 11" or A4
paper, in an easy-to-use, portable device.
High-Resolution Color LCD
High-resolution color display provides real-time preview of
12-lead ECG and post-acquisition review of acquired ECG.
Distinguished Mortara VERITAS™ Resting
ECG Interpretation Algorithm
Widely recognized Mortara resting ECG interpretation
algorithm uses gender specific and adult and pediatric
criteria to provide a silent second opinion for resting
ECG interpretation.
Choice of Wireless or Traditional
ECG Acquisition
ELI 250c offers a choice of either the innovative WAM™
wireless acquisition module or the AM12™ acquisition
module. Both include replaceable lead wires, lead fail
indicator, and remote control with buttons for ECG
acquisition and rhythm printing.
Best 10
ELI 250c automatically selects and displays the Best 10
seconds of ECG signal from the five minute segment of
internal full disclosure, reducing clinical review time and
eliminating the need for repeat ECGs.
Full Keyboard
Alphanumeric elastomer keyboard features dedicated
“one-touch” buttons for ECG acquisition, rhythm printing,
and ECG transmission/order retrieval.
Information Exchange
Experience the benefits of bidirectional communication
via internal modem, LAN, or wireless LAN with Mortara’s
E-Scribe™ and Athena product, as well as third party
EMR, HIS, and PACS systems via XML, PDF, UNIPRO32,
and DICOM®32.
Technical Specification
INSTRUMENT TYPE: 12-lead electrocardiograph
INPUT CHANNELS: Simultaneous acquisition of all 12 leads
STANDARD LEADS ACQUIRED: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
WAVEFORM DISPLAY: Backlit, 1/4 VGA color LCD (320 x 240); 3, 4+4, or 6+6 lead presentation
INPUT IMPEDANCE: Meets or exceeds the requirements of ANSI/AAMI EC11
INPUT DYNAMIC RANGE
ELECTRODE OFFSET TOLERANCE
COMMON MODE REJECTION
PATIENT LEAKAGE CURRENT:
Meets or exceeds the requirements of ANSI/AAMI ES1
CHASSIS LEAKAGE CURRENT
DIGITAL SAMPLING RATE:
40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
OPTIONAL FUNCTIONS:
Mortara VERITAS resting ECG interpretation algorithm with age and gender specific criteria; connectivity with bidirectional communication
PAPER: Perforated Z-fold thermal paper, A4 or 8.5 x 11" wide, 250 sheets
THERMAL PRINTER: Computer-controlled dot array; 8 dots/mm
THERMAL PRINTER SPEEDS: 5, 10, 25, or 50 mm/s
GAIN SETTINGS: 5, 10, or 20 mm/mV
REPORT PRINT FORMATS: Standard or Cabrera: 3+1, 3+3, 6, 6+6, or 12 channel
RHYTHM PRINT FORMATS: 3, 6, or 12 channel with configurable lead groups
KEYBOARD:
Elastomer keyboard with complete alphanumeric keys, soft-key menu, and dedicated function keys
FREQUENCY RESPONSE: 0.05 to 300 Hz
FILTERS:
High-performance baseline filter; AC interference filter
50/60 Hz; low-pass filters 40 Hz, 150 Hz, or 300 Hz
A/D CONVERSION: 20 bits (1.17 microvolt LSB)
DEVICE CLASSIFICATION: Class I, Type CF defibrillation-proof applied parts
ECG STORAGE: Internal storage up to 100 ECGs; optional expanded up to 200 ECGs
WEIGHT: 11.25 lbs. (5.1 kg) including battery (without paper)
DIMENSIONS: 15.5 x 17 x 4" (39.4 x 43.2 x 10.2 cm)
POWER REQUIREMENTS:
Universal AC power supply (100-240 VAC at 50/60 Hz)
110 VA; internal rechargeable battery